3 Phase Power Quality Analyzers
Products Support About Us Contact
PQProbe Specifications     
Input Characteristics

Voltage inputs
Number of inputs 4(3 phases + neutral)
Maximum input voltage 707 Vrms
Nominal voltage range Selectable from 1V to 600V. IEC 61000-4-30 compliant
Maximum peak measurement voltage 1 kV
Bandwidth > 10 kHz, up to 100 kHz for transient display
Nominal Frequency 50 Hz/60 Hz
Scaling User Programmable
Current inputs
Number of inputs 4 (3 phases + neutral)
Type Clamp on current transformer with mV output
Range 1 Arms to 3000 Arms with appropriate clamps
Bandwidth upto 10 kHz
Nominal Frequency 40 Hz to 70 Hz
Scaling User Programmable
Sampling system
Resolution 13 bit analog to digital converter on 8 channels
Sampling Speed 200 kS/s on each channel simultaneously
PLL synchronization channel source Voltage for channel A, B or C is scanned for PLL synchronization
PLL synchronization frequency range 42.5 Hz to 69 Hz

Standards

Measurement methods used IEC 61000-4-30 Class-A
Measurement performance IEC 61000-4-30 Class-A
Power quality EN 50160
Flicker IEC 61000-4-15
Harmonics IEC 61000-4-7

Accuracy, resolution and range

Measurement range Resolution Accuracy
Volt/Amps/Power Factor/Hertz
Vrms 1 Vrms to 600 Vrms 0.01 Vrms ± 0.1% of nominal voltage
Vpk 1 Vpk to 1000Vpk 1 V 5 % of nominal voltage
Voltage Crest Factor 3 or less 0.01 ± 5%
Arms 0 Arms to 3000 Arms 1 Arms ± 0.1% ± 10 counts1
Apk using 1 mV/A scaling 0 Apk to 5250 Apk 1 A ± 5%
A Crest Factor 4 or less 0.01 ± 5%
Power Factor 0 to 1 0.01 ± 0.033
Displacement Power Factor 0 to 1 0.01 ± 0.033
Hz @ 50 Hz nominal 42.500 Hz to 57.500 Hz 0.001 Hz ± 0.01 Hz
Hz @ 60 Hz nominal 51.000 Hz to 69.000 Hz 0.001 Hz ± 0.01 Hz
Harmonics
Harmonic Order 1 to 50 grouping, harmonic groups according to IEC 61000-4-7
Inter-Harmonic order 1 to 49 grouping, inter-harmonic subgroups according to IEC 61000-4-7
Sub-Harmonic order @ 50 Hz nominal 10 Hz to 45 Hz accumulated subgroup according to IEC 61000-4-7
Sub-Harmonic order @ 60 Hz nominal 10 Hz to 55 Hz accumulated subgroup according to IEC 61000-4-7
Above-Harmonic order @ 50 Hz nominal 2100 Hz to 8900 Hz, above-harmonic subgroups according to IEC 61000-4-7
Above-Harmonic order @ 60 Hz nominal 2100 Hz to 8880 Hz, above-harmonic subgroups according to IEC 61000-4-7
Vrms Magnitude 0 Vrms to 600 Vrms 0.1 Vrms ± 0.05% of nominal voltage if < 1% nominal voltage.
± 5% if ≥ 1% nominal voltage
Arms Magnitude 0 Arms to 3000 Arms 1 Arms ± 0.15% of sliding reference current if < 3% sliding reference current.
± 5% if ≥ 3% sliding reference current
Watt Magnitude Depends on clamp and voltage scaling ± 1.5 W if < 150 W, ± 1% of nominal power if ≥ 150 W
THD (50) Relative 0.0% to 100.0% 0.1% ± 2.5% V and A (± 5% Watt)
Phase Angle -360° to +0° ± n X 1°
Flicker
Pst (1min), Pst, Plt instantaneous Flicker 0.00 to 20.00 0.01 Within ± 5 % of tabulated values according IEC61000-4-15
Power and energy
Watt [VA, VAR] Depends on clamp and voltage scaling ± 1% ± 10 counts
kWh [kVA, kVAR] Depends on clamp and voltage scaling ± 1% ± 10 counts
Imbalance
Volts [neg. and zero seq.] 0.0 % to 5.0 % 0.1% ± 0.15 %
Current [neg. and zero seq.] 0.0 % to 20 % 0.1% ± 1 %

Measurement Modes

Standard Power Quality, Demand, Energy This mode collects data for troubleshooting using standard power quality parameters with event conditions set.
Current Inrush Study This mode records motor start-up and inrush conditions. This mode is used when recording is triggered on change in current.
Fault Recorder Study This mode is similar to current Inrush but with a voltage trigger.
Long Term Timed Recording, Demand, Energy This mode is used for long term statistical studies and benchmarking field-based equipment testing and evaluation. No triggers are set so only trended values are available.
Continuous Data Logging, Demand, Energy This mode logs rms and power values once per second.
EN50160 Power Quality Study This mode monitors and reports Quality of Supply (QoS) compliance as specified by the EN50160 standard.

Measurement method

Vrms, Arms 10/12 cycle contiguous non overlapping intervals in accordance with IEC 61000-4-30
Vpeak, Apeak Absolute highest sample value within 10/12 cycle interval
V Crest Factor Measures ratio between the Vpeak and Vrms
A Crest Factor Measures ratio between the Apeak and Arms
Hz Measured every 10 sec in accordance with IEC61000-4-30
Vrms˝, Arms˝ Value is measured over 1 cycle, commencing at a fundamental zero crossing, and refreshed each half-cycle. This technique is independent for each channel in accordance with IEC 61000-4-30.
Harmonics Calculated from 10/12-cycle gapless harmonic group measurements on Voltage and Amps according to IEC 61000-4-7.
Watt Calculates average value of instantaneous power over 10/12 cycle period for each phase. Total Active Power PT = P1 + P2 + P3.
VA Calculates apparent power using Vrms x Arms value over 10/12 cycle period. Total apparent power is root mean square of real and reactive power.
VAR Calculates reactive power as root of VA squared minus watt squared over 10/12 cycle period.
Power Factor Calculated watt/VA.
Displacement Power Factor Cos of angle between fundamental voltage and current.
Imbalance The supply voltage/current imbalance is evaluated using the method of symmetrical components according to IEC61000-4-30. Imbalance is also calculated according to NEMA guidelines.

Events

Measurement range Resolution Accuracy
Voltage Magnitude (Sags, Swells and Interruption) Event
Vrms˝ 0.0% to 200.0% of nominal voltage 0.1 Vrms ± 0.2% of nominal voltage
Threshold Levels Programmable thresholds in percent of nominal voltage. Event detection based upon ˝ cycle rms voltages. Captures sags, swells and interruptions voltage changes
Duration hhh, mm, sss, mmm Half cycle One cycle
Transient Event
Volts rms reading 10 Vrms to 707 Vrms 1 V ± 2.5% of nominal
Minimum detect duration 5 µs
Sampling rate 200 kS/s
Current Inrush Event
Arms˝ 0 Arms to 3000 Arms 1 Arms ± 0.5% ± 10 counts
Threshold Levels Programmable thresholds in percent of sliding reference value of current. Event detection based upon ˝ cycle rms current. Captures rapid current changes.
Duration hhh, mm, sss, mmm Half cycle One cycle
Frequency Event
Frequency Programmable thresholds in percent of nominal frequency. Event detection based upon 10/12 cycle calculated value.

Software Display Modes

Meter Presents measurements of last calculation period, phasor diagrams, scope graphs, harmonic plots, and events summary on a single screen. The screen can be customized to display information as required by user.
Trend Plots graphs for continuous data logging measurement performed. The parameters displayed can be changed as per user inputs. The min/max/average values of any parameter at any point are readily available using the mouse cursor.
Event Presents listing of all events within a continuous measurement in the order they occurred. Each record contains date/time when event occurred, event category, channel and duration of event and characterization details.

Wiring Configurations

Single Phase Single Phase with neutral
1Ø Split Phase Split Phase
3Ø WYE Three phase four wire system WYE
3Ø DELTA Three phase three wire system DELTA
3Ø Open Leg Open delta three wire system with 2 Watt
Generic Any other wiring configuration

Environment and Safety Specifications

Operating Environment Indoors, up to 2000 m (6562-ft.) ASL

Temperature and Humidity

Storage -10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
Operation 0 to 40°C (32 to 104°F), 80% RH or less (non-condensating)

 








At Enetranet, our mission is to provide simple to use technology solutions, excellent product support and power quality expertise to commercial and industrial (C&I) customers , electrical utilities and electrical consultants to help them diagnose power quality problems thereby reduce electrical operating costs, improve system reliability and efficiency.